Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Theory of Operation 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
206 
9.3.5.3. Zero/Span Valve Options 
Any zero/span valve options installed in the analyzer are controlled by a set of electronic switches located 
on the relay board.  These switches, under CPU control, supply the +12VDC needed to activate each 
valve’s solenoid. 
9.3.5.4. IR Source 
The Relay board supplies a constant 11.5VDC to the IR Source.  Under normal operation the IR source is 
always on.   
9.3.5.5. Status LEDs 
Eight LEDs are located on the analyzer’s relay board to show the current status on the various control 
functions performed by the relay board (see Figure 10-14).  They are: 
Table 9-2:     Relay Board Status LED’s 
LED 
COLOR 
FUNCTION 
STATUS WHEN LIT 
STATUS WHEN UNLIT 
D1 RED 
Watchdog 
Circuit 
Cycles On/Off Every 3 Seconds under direct control of the 
analyzer’s CPU. 
D2 
YELLOW 
Wheel Heater 
HEATING 
NOT HEATING 
D3 
YELLOW 
Bench Heater 
HEATING 
NOT HEATING 
D4 YELLOW 
Spare 
N/A 
N/A 
D5 GREEN Sample/Cal Gas 
Valve Option 
Valve Open to CAL GAS 
FLOW 
Valve Open to SAMPLE GAS 
FLOW 
D6 GREEN Zero/Span Gas 
Valve Option 
Valve Open to SPAN GAS 
FLOW 
Valve Open to ZERO GAS FLOW 
D7 GREEN  Shutoff Valve 
Option 
Valve Open to CAL GAS 
FLOW 
Valve CLOSED to CAL GAS 
FLOW 
D8 
GREEN 
IR SOURCE 
Source ON 
Source OFF