Theory of Operation 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
211 
9.3.8. Power Supply/ Circuit Breaker 
The analyzer operates on 100 VAC, 115 VAC or 230 VAC power at either 50Hz or 60Hz.  Individual units 
are set up at the factory to accept any combination of these five attributes.  As illustrated in Figure 10-15, 
power enters the analyzer through a standard IEC 320 power receptacle located on the rear panel of the 
instrument.  From there it is routed through the On/Off switch located in the lower right corner of the Front 
Panel.  A 6.75 Amp circuit breaker is built into the ON/OFF Switch. 
AC power is distributed directly to the sample gas pump.  The bench and GFC wheel heaters as well as 
the GFC wheel receive AC power via the relay board.   
AC Line power is converted stepped down and converted to DC power by two DC power supplies.  One 
supplies +12 VDC, for valves and the IR source, while a second supply provides +5 VDC and ±15 VDC 
for logic and analog circuitry.  All DC voltages are distributed via the relay board. 
 
 
CAUTION 
Should the AC power circuit breaker trip, investigate and correct the condition causing 
this situation before turning the analyzer back on.
 
 
 
AC POWER 
ENTRANCE 
ON/OFF 
SWITCH 
Cooling Fan 
Mother  
Board 
CPU 
 
PS 1  (+5 VDC; ±15  VDC) 
PS 2 (+12 VDC) 
Display 
LVDS 
transmittedr board 
Pressure 
Sensors 
IR Source 
Pump
GFC   Wheel 
Motor 
Sync/Demod 
M/R & 
Segment 
Sensors 
Heaters 
Valve Options 
KEY 
AC POWER 
DC POWER 
 
 
 
RELAY  
BOARD 
US
B
 
Touchscreen 
 
Figure 9-16: 
Power Distribution Block Diagram