Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Specifications 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
GFC 7000TA Parameter 
GFC 7000TA Specification 
Weight  
40 lbs.  (18.1 kg) 
Certifications 
CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class A 
1
 At constant temperature and voltage. 
 
Table 2-2:  Model GFC 7000TM Basic Unit Specifications 
GFC 7000TM Parameter 
GFC 7000TM Specification 
Ranges  
(Physical Analog Output) 
Min: 0-4 ppm Full scale 
Max: 0-4000 ppm Full scale
 
 
Selectable, dual ranges and auto ranging supported 
Measurement Units 
ppb, ppm, µg/m
3
, mg/m
3
, (selectable) 
Zero Noise 
< 0.2 ppm (RMS) 
Span Noise 
< 1% of reading (RMS) 
Lower Detectable Limit 
< 0.4 ppm
1
 
Zero Drift (24 hours) 
<0.5 ppm
1
 
Span Drift (24 hours) 
<0.5% of reading
1
 
Lag Time 
10 seconds 
Rise/Fall Time  
<60 seconds to 95%  
Linearity 
1% of full scale 
Precision 
0.5% of reading  
Sample Flow Rate 
800cm
3
/min.  ±10% 
O
2
 Sensor option adds 110 cm³/min, ±20%, to total flow though when installed 
Temperature Coefficient 
< 0.1% of Full Scale per 
o
C or 0.1 ppm per 
o
C, whichever is greater 
Voltage Coefficient 
< 0.05% of Full Scale per V 
AC Power Requirements 
100V – 120V, 220V – 240V, 50/60 Hz  
Analog Output Ranges 
10V, 5V, 1V, 0.1V  
Recorder Offset 
±10% 
Analog Output Resolution 
1 part in 4096 of selected full-scale voltage 
Standard I/O 
1 Ethernet: 10/100Base-T 
2 RS232 (300-115,200 baud) 
2 USB device ports 
8 Status opto-isolated digital status outputs 
6 Opto-isolated digital control inputs (3 defined, 3 spare) 
4 Analog outputs 
Optional I/O 
1 USB com port 
1 RS485 
8 Analog inputs (0-10V, 12-bit) 
4 Digital alarm outputs 
Multidrop RS232 
3 4-20mA current outputs 
Operating Temperature Range 
5-40
o
Humidity Range 
0 - 95% RH, non-condensing 
Dimensions H x W x D 
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm) 
Weight  
40 lbs.  (18.1 kg) 
Environmental Installation 
category (over-voltage category) II; Pollution degree 2