Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
Teledyne Analytical Instruments 
3. GETTING STARTED 
3.1. Unpacking and Initial Set Up 
 
CAUTION 
To avoid personal injury, always use two persons to lift and carry the Model GFC 
7000TA.
 
 
 
WARNING 
Never disconnect PCAs, wiring harnesses or electronic subassemblies while the 
instrument is under power.
 
 
 
NOTE
 
It is recommended that you store shipping containers/materials for future use if/when the instrument should 
be returned to the factory for repair and/or calibration service. See Warranty section in this manual and 
shipping procedures on our
 Website at: 
http://www.teledyne-api.com
 under Customer Support > Return Authorization.
 
 
1.  Verify that there is no apparent external shipping damage.  If damage has occurred, please 
advise the shipper first, then Teledyne Instruments. 
2.  Included with your analyzer is a printed record of the final performance characterization 
performed on your instrument at the factory.  This record, titled Final Test and Validation Data 
Sheet
 (P/N 04596) is an important quality assurance and calibration record for this instrument.  It 
should be placed in the quality records file for this instrument. 
3.  Carefully remove the top cover of the analyzer and check for internal shipping damage. 
  Remove the set-screw located in the top, center of the Front panel. 
  Remove the 2 screws fastening the top cover to the unit (one per side towards the rear).   
  Slide the cover backwards until it clears the analyzer’s front bezel. 
Lift the cover straight up. 
 
 
 
 
CAUTION – Avoid Warranty Invalidation 
Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too 
small to be felt by the human nervous system.  Damage resulting from failure to 
use ESD protection when working with electronic assemblies will void the 
instrument warranty.   
See  A Primer on Electro-Static Discharge in this manual for more information on 
preventing ESD damage.