Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
16 
Table 3-2:     Rear Panel Description 
Component Function 
cooling fan 
Pulls ambient air into chassis through side vents and exhausts through rear  
 
AC power 
connector 
Connector for three-prong cord to apply AC power to the analyzer  
CAUTION! The cord’s power specifications (specs) MUST comply with the power 
specs on the analyzer’s rear panel Model number label  
Model/specs label 
Identifies the analyzer model number and provides power specs 
PURGE IN 
 
SAMPLE 
Connect a gas line from the source of sample gas here   
Calibration gases are also inlet here on units without zero/span valve options installed 
EXHAUST 
Connect an exhaust gas line of not more than 10 meters long here that leads outside 
the shelter or immediate area surrounding the instrument 
VENT SPAN 
Span gas vent outlet for units with zero/span valve options installed 
Connect an exhaust gas line of not more than 10 meters long here 
PRESSURE SPAN 
On units with zero/span valve options installed, connect a gas line to the source of 
calibrated span gas here 
IZS 
Internal Zero Air: On units with zero/span valve options installed but no internal zero 
air scrubber attach a gas line to the source of zero air here 
RX TX 
LEDs indicate receive (RX) and transmit (TX) activity on the when blinking 
COM 2 
Serial communications port for RS-232 or RS-485 
RS-232 
Serial communications port for RS-232 only 
DCE DTE 
Switch to select either data terminal equipment or data communication equipment 
during RS-232 communication  
STATUS 
For outputs to devices such as Programmable Logic Controllers (PLCs)  
ANALOG OUT 
For voltage or current loop outputs to a strip chart recorder and/or a data logger 
CONTROL IN 
For remotely activating the zero and span calibration modes  
ALARM 
Option for concentration alarms and system warnings  
ETHERNET 
Connector for network or Internet remote communication, using Ethernet cable  
ANALOG IN 
Option for external voltage signals from other instrumentation and for logging these 
signals 
USB  
Com port optional connector for direct connection to laptop computer, using USB cable 
Information Label 
Includes voltage and frequency specifications