Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
28 
 
7.  Attach a sample inlet line to the sample inlet port.  The SAMPLE input line should not be more 
than 2 meters long. 
NOTE 
Ideally, the pressure of the sample gas should be at ambient pressure (0 psig).  
Maximum pressure of sample gas should not exceed 1.5 in-Hg over ambient. 
In applications where the sample gas is received from a pressurized manifold, a vent 
must be placed as shown to equalize the sample gas with ambient atmospheric 
pressure before it enters the analyzer. 
This vent line must be: 
  At least 0.2m long 
  No more than 2m long and vented outside the shelter or immediate area 
surrounding the instrument.
 
 
8.  Attach sources of zero air and span gas (see Figures 3-3 through 3-8 inclusive). 
  Span Gas is a gas specifically mixed to match the chemical composition of the type of gas 
being measured at near full scale of the desired measurement range.   
In the case of CO
2
 measurements made with the Teledyne Instruments Model GFC 7000TA 
Analyzer it is recommended that you use a gas calibrated to have a CO
2
 content equaling 
80% of the range of compositions being measured. 
EXAMPLE:  If the application is to measure between 0 ppm and 500 ppm, an appropriate Span 
Gas would be 400 ppm.  If the application is to measure between 0 ppm and 100 ppm, an 
appropriate Span Gas would be 80 ppm.   
  Span Gas can be purchased in pressurized canisters or created using Dynamic Dilution 
Calibrator such as the Teledyne Instruments Model T700 and a source of dried air scrubbed 
of CO
2
 such as a bottle of N
2
  Zero Air is similar in chemical composition to the Earth’s atmosphere but scrubbed of all 
components that might affect the analyzer’s readings.   
In the case of CO
2
 measurements this means less than 0.1 ppm of CO
2
 and Water Vapor 
(when dew point <-15˚C).  Zero Air can be purchased in pressurized canisters. 
9.  Attach an exhaust line to the exhaust outlet port.   
  The exhaust from the pump and vent lines should be vented to atmospheric pressure using 
maximum of 10 meters of ¼” PTEF tubing.   
 
CAUTION 
Venting should be outside the shelter or immediate area surrounding the instrument.