Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
33 
6.  Turn the instrument on. 
7.  Make sure the zero air supply to the analyzer’s purge line inlet is supplying gas at a stable 
pressure above 7.5 psig. 
8.  Adjust the GFC 7000TA’s pressure regulator until the attached gauge reads 7.5 psig. 
9.  Turn off the instrument.   
10.  Remove the source of zero air attached to the purge line inlet port at the back of the analyzer. 
11.  Remove the pressure gauge and reattach the end cap removed in step 4 above. 
12.  Replace the analyzer’s top cover.   
3.7. Initial Operation 
If you are unfamiliar with the GFC 7000TA theory of operation, we recommend that you read  
Section 10.  For information on navigating the analyzer’s software menus, see the menu trees described 
in Appendix A. 
NOTE 
The analyzer’s cover must be installed to ensure that the temperatures of the GFC 
wheel and absorption cell assemblies are properly controlled. 
3.7.1. Startup 
After electrical and pneumatic connections are made, and initial functional check is required. Turn on the 
instrument.  The pump, exhaust fan and PMT cooler fan should start immediately.  The display will briefly 
show a logo splash screen at the start of initialization. 
The analyzer should automatically switch to Sample Mode after completing the boot-up sequence and 
start monitoring CO
2
 gas. However, there is an approximately one hour warm-up period before reliable 
gas measurements can be taken. During the warm-up period, the front panel display may show 
messages in the Parameters field.