 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
56 
 
The minimum span gas flow rate required for this option is 800 cm
3
/min.   
The state of the zero/span valves can also be controlled: 
  Manually from the analyzer’s front panel by using the SIGNAL I/O controls located under the DIAG 
Menu (Section 6.13.2), 
  By activating the instrument’s AutoCal feature (Section 7.6), 
  Remotely by using the external digital control inputs (Sections 6.15.1.2 and 7.5.2), or 
  Remotely through the RS-232/485 serial I/O ports (see Appendix A-6). 
 
GFC Wh
ee
l
Hous
ing
          
SAMPLE 
CHAMBER
GFC M
o
tor
 
He
at Sy
nc
Purg
e Gas
F
low
 C
o
nt
ro
l
Ext
e
rn
al CO2
 
Sc
rub
ber 
(O
pt
io
n
 53
 0n
ly
)
 
Figure 5-3: 
Internal Pneumatic Flow – Ambient Zero/Ambient Span