Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
75 
TEMP: 
Samples the analyzer’s bench temperature, box temperature and PHT cooler drive voltage every 
five minutes and records an average once every six hours.  Data from the last 400 averaging periods is 
recorded.  A time and date stamp is recorded for every data point logged.  This information is useful for 
performing predictive diagnostics as part of a regular maintenance schedule (See Section 9.1). 
Note 
The CALDAT, STBZRO and STBSPN channels collect data based on events (e.g.  a 
calibration operation) rather than a timed interval.  This does not represent any 
specific length of time since it is dependent on how often calibrations are 
performed. 
 
Triggering Events and Data Parameters/Functions for these default channels are: 
 
LIST OF CHANNELS
LIST OF PARAMETERS
NAME: CONC 
EVENT: ATIMER 
REPORT PERIOD: 000:01:00 
NO. OF RECORDS: 800 
RS-232 REPORT: OFF 
CHANNEL ENABLED: ON 
CAL HOLD OFF: ON
 
PARAMETER: CONC1 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
NAME: PNUMTC
EVENT: ATIMER
REPORT PERIOD: 001:00:00
NO. OF RECORDS: 360
RS-232 REPORT: OFF
CHANNEL ENABLED: ON
CAL HOLD OFF: OFF
PARAMETER: SMPLFLW 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: SMPLPRS 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
NAME: CALDAT 
EVENT: SLPCHG 
REPORT PERIOD: N/A 
NO. OF RECORDS:200 
RS-232 REPORT: OFF 
CHANNEL ENABLED: ON 
CAL HOLD OFF: OFF
 
PARAMETER: OFSET1 
MODE: INST 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: SLOPE1 
MODE: INST 
PRECISION:
STORE NUM SAMPLES OFF
 
PARAMETER: ZSCNC1 
MODE: INST 
PRECISION: 
STORE NUM SAMPLES OFF
 
NAME: STBZRO
EVENT: EXITZR
REPORT PERIOD: N/A
NO. OF RECORDS:200
RS-232 REPORT: OFF
CHANNEL ENABLED: ON
CAL HOLD OFF: OFF
PARAMETER: DETMES 
MODE: INST 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: STABIL 
MODE: INST 
PRECISION:
STORE NUM SAMPLES OFF
 
PARAMETER: RATIO 
MODE: INST 
PRECISION: 
STORE NUM SAMPLES OFF
 
NAME: STBSPN 
EVENT: EXITSP 
REPORT PERIOD: N/A 
NO. OF RECORDS:200 
RS-232 REPORT: OFF 
CHANNEL ENABLED: ON 
CAL HOLD OFF: OFF
 
PARAMETER: DETMES 
MODE: INST 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: RATIO 
MODE: INST 
PRECISION: 
STORE NUM SAMPLES OFF
 
NAME: TEMP
EVENT: ATIMER
REPORT PERIOD: 000:06:00
NO. OF RECORDS:400
RS-232 REPORT: OFF
CHANNEL ENABLED: ON
CAL HOLD OFF: OFF
PARAMETER: BOXTMP 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
PARAMETER: BNTEMP 
MODE: AVG 
PRECISION:
STORE NUM SAMPLES OFF
 
PARAMETER: PHTDRV 
MODE: AVG 
PRECISION: 
STORE NUM SAMPLES OFF
 
LIST OF PARAMETERS 
 
Figure 6-4: 
Default DAS Channels Setup