Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
78 
6.7.2.3. Editing DAS Data Channels 
DAS configuration is most conveniently done through the APICOM remote control program.  The 
following list of key strokes shows how to edit using the front panel. 
 
 
 
 
 
Main Data Acquisition Menu 
Edit Data Channel Menu
SAMPLE*          RANGE = 500.000 PPM       CO2 =X.XXX   
 
 < TST  TST >  CAL  
 
 
SETUP
 
 
 
 
SETUP X.X     
  0) CONC:     ATIMER,     1,          900
 
 
 
PREV  NEXT
    
             INS      DEL     EDIT         PRNT       EXIT
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG   
DAS
  
RNGE
  
PASS  CLK  MORE 
EXIT
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
 
EXIT
 will return to the 
previous SAMPLE 
display. 
SETUP X.X       
     
DATA ACQUISITION
 
 
 
 
 VIEW
 
 
 
 
 
 
EDIT
 
 
 
 
 
EXIT
 
Moves the 
display up & 
down the list of 
Data Channels 
Inserts a new Data 
Channel into  the list 
BEFORE the Channel 
currently being displayed  
Deletes The Data 
Channel currently 
being displayed 
Exports the 
configuration of  all 
data channels to 
RS-232 interface. 
Exits to  the Main 
Data Acquisition 
Menu 
 
 
 
SETUP X.X      
NAME:CONC
 
<SET SET>
 
 EDIT  PRNT       
 
EXIT
 
Moves the display 
between the 
PROPERTIES for this 
data channel. 
Reports the configuration of current 
data channels to the RS-232 ports. 
Exits returns to the 
previous Menu 
Allows to edit the channel name, see next key sequence. 
 
When editing the data channels, the top line of the display indicates some of the configuration 
parameters.  For example, the display line: 
0) CONC: ATIMER, 4, 800
 
translates to the following configuration: 
Channel No.: 0 
NAME: CONC 
TRIGGER EVENT: ATIMER 
PARAMETERS: Four parameters are included in this channel 
EVENT: This channel is set up to record 800 data points.