Model 4501-XX Toxic Gas Sensor Module
 
6.3 DIAGNOSTICS SUB-MENU 
 
Function
Display
Description
Reference
M
E
S T
M
E
S T
M
E
S T
M
E
S T
M
E
S T
M E
S
T
M
E
S T
M
E
S T
Table 6-1
Calibration Sub-Menu
Sub-Routine A - Abort Calibration
H2S
XXX PPM
Default Display
H2S
CAL-FAIL
Calibration Failed
Sub-Routine C - Calibration Fail
Enter
READ
MANUAL
Operation:  Remove gas from sensor read manual, particularly section 6.3 and 
7.6, to check calibration procedures.
H2S
CAL-OK
Calibration successful
Mode
CALIB
ABORTED
Operation:  Hold magnet over  [M]  for 10 seconds to abort calibration
Sub-Routine B - Select Span
25PPM SPAN
Banner: [S] or [T] to change span value, [E] when done
Down
C##PPM
Banner: Apply xx-PPM gas, [E] when stable
Should be stable after 3 
minutes of gas exposure
Enter
H2S
REMOVE GAS
Operation: Remove gas from sensor
If calibration failed, then 
Sub-Routine C
Enter
H2S
CAL-OK
Banner: Remove gas from sensor
Sub-Routine C
ZERO
DONE
Zero gas setting completed (Message will appear on the screen and after 3 
seconds will go to the next menu - 25PPM Span)
Operation:  Confirm area clear of gas, or apply  zero air to sensor.
Enter
CAL 0 PPM
Banner:  Apply zero gas, [E] when done
or Sub-Routine A
Key
Mode
H2S
CALIB:
Default Display (example 4501-05)
XXPPM
Banner: [S] or [T] to change span value, [E] when done
Sub-Routine B
Enter
Cal0PPM
Wait.XXX
There is a minimum time to calibrate at 0PPM. If [E] is hit prior to this time 
has elapsed, a countdown screen will appear.
H2S
WAIT-148
Operation:  148 second time out before sensor is returned to service.
 
Page:  14