Using Dicom Theater software
83
Stopping and starting stabilization
Introduction
Stabilization starts automatically after the projector was 
calibrated from MediCal Pro. 
The stabilization process can be stopped and restarted explicitly 
by the user using the Dicom Theater application. 
However, if the stabilization process terminates automatically, 
then the user cannot restart it again via the application. The user 
is required to go to MediCal Pro and activate the preset of her or 
his choice. This will re-activate the stabilization process. 
E.g., in case the user switches inputs or presets by means of the 
projector’s remote control unit, stabilization will stop 
automatically. It can be re-started only by activating a preset in 
MediCal Pro.
Stopping stabilization manually
If for some reason you wish to stop the projector stabilization 
(e.g., in case there is too much ambient light), proceed as 
follows:
1. Position the cursor over the Dicom Theater sys-tray icon and 
click the right mouse button (“right-click”).
The Dicom Theater menu appears.
Figure 32
2. Click Stop Stabilizer.
The stabilization process will stop.