Справочник Пользователя для HP (Hewlett-Packard) 22A

Скачать
Страница из 508
Rack Flange Kit for
Option 913 synthesizers are supplied with rack flanges and the
Synthesizers with
necessary hardware to install them on the synthesizer without
Handles Attached
removing the instrument handles. The following table itemizes the
(Option 913)
parts in this kit.
Table 3-5.
Rack Flange Kit for Synthesizers with Handles Attached
Contents
CAUTION
Ventilation Requirements: 
When installing the instrument in a
cabinet, the convection into and out of the instrument must not be
restricted. The ambient temperature (outside the cabinet) must be
less than the maximum operating temperature of the instrument
by 4 
 for every 100 watts dissipated in the cabinet. If the total
power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
INSTALLATION 3-15