MODEL 5081-P pH/ORP
SECTION 10.0
PROGRAMMING  FOR ORP MEASUREMENTS
10.3.3 Procedure
1. Enter  the  Program  menu  by  pressing    PROG    on  the  IRC.  The  OutPut sub-menu
appears.
2. Press ENTER. The screen displays the 4 MA prompt. use the editing keys to
change the displayed number to the desired ORP. The allowed range is -1400 to
1400. To change the display to a negative number, press è or  ç until no digit is
flashing. Then press é or ê until the minus sign appears. To change the display to
a positive number, press è or  ç until the negative sign is flashing. Then press
é
or ê until the minus sign disappears. Press ENTER  to save.
3. The screen displays the 20 MA prompt. use the editing keys to change the displayed
number to the desired ORP. The allowed range is -1400 to 1400. Press  ENTER  to
save.
4. The screen displays the HoLd prompt. use the editing keys to change the display to
the  output  desired  when  the  transmitter  is  in  hold. The  range  is  3.80  to  22.00  mA.
Entering  00.00  causes  the  transmitter  to  hold  the  output  at  the  value  it  was  when
placed in hold. The hold setting overrides the fault setting. Press  ENTER  to save.
5. The screen displays the FAULt prompt. use the editing keys to change the display to
the output desired when the transmitter detects a fault. The range is 3.80 to 22.00 mA.
Entering 00.00 causes the transmitter to hold the output at the value it was when the
fault occurred. Press  ENTER  to save.
6. The screen displays the dPn prompt. use the editing keys to change the display to the
desired output dampening. The range is 0 to 255. Press  ENTER  to save.
7. Press  EXIT  to return to the process display. 
PROGRAM
HoLd
EXIT
NEXT       ENTER
2
1
.00
PROGRAM
FAULt
EXIT
NEXT       ENTER
2
2
.00
PROGRAM
dPn
EXIT
NEXT       ENTER
PROGRAM
OutPut
EXIT
NEXT       ENTER
PROGRAM
4MA
EXIT
NEXT       ENTER
-
1400
PROGRAM
20MA
EXIT
NEXT       ENTER
1
400
0
00