MODEL 5081-P pH/ORP
SECTION 10.0
PROGRAMMING  FOR ORP MEASUREMENTS
PROGRAM
rEF
EXIT
ENTER
LO
PROGRAM
IMPtC
EXIT
NEXT       ENTER
ON
PROGRAM
rOFFSt
EXIT
ENTER
PROGRAM
dIAGnOSTIC
EXIT
NEXT       ENTER
0
60
PROGRAM
dIAG
EXIT
ENTER
OFF
10.6.3 Procedure
1. Press PROG on the infrared remote controller (IRC).
2. Press NEXT until the diAGnOStIC sub-menu appears. Press  ENTER.
3. The screen displays the rOFFSt prompt. use the editing keys to change the flashing
display to the desired standardization (reference) offset (in millivolts). The range is 0 to
1000 mV. Press ENTER to save.
4.  The dIAG prompt appears. use the é or  ê keys to enable (On) or disable (OFF) the
sensor diagnostics. Press ENTER to save.
5. The IMPtC prompt appears. use the é or  ê keys to enable (On) or disable (OFF)
glass impedance temperature compensation. Because glass impedance is a strong
function of temperature, correcting glass impedance for temperature is recommended. A
third setting (SPC) appears in addition to On and OFF. do not select SPC; the setting is
intended for factory use. Press ENTER to save.
6. The rEF prompt appears. Press é or  ê until the desired setting appears. LO identifies
a low impedance reference electrode, and HI identifies a high impedance reference
electrode. Press ENTER to save. Selecting LO disables the low impedance warning and
failure limits for the reference electrode.
7. The rFH prompt appears. use the editing keys to change the display to the desired ref-
erence electrode high impedance fault value. The allowed ranges are
Type of reference electrode
Allowed range
Low impedance (LO in step 6)
0 - 2000 kilohms
High impedance (HI in step 6)
0 - 2000 megohms
Entering 0000 disables the feature. When the reference electrode impedance goes
above the fault value, the transmitter displays the diagnostic message rEFFAIL and sets
a fault condition. Press ENTER  to save.
PROGRAM
rFH
EXIT
ENTER
1