81
MODEL 5081-P pH/ORP
SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
4. REFERENCE IMPEdANCE. The majori-
ty of reference electrodes used in indus-
try are low impedance silver-silver chlo-
ride electrodes. Every pH and ORP sen-
sor  manufactured  by  Rosemount
Analytical  has  a  low  impedance  refer-
ence.  However,  there  are  applications
that call for either a high impedance sodi-
um or pH glass reference electrode. Both
high impedance and low impedance ref-
erence electrodes can be used with the
Model 5081-P pH/ORP transmitter.
5. WARNING ANd FAILuRE LIMITS FOR
THE  REFERENCE  ELECTROdE.
Warning tells the user that the reference
electrode impedance is approaching the
failure limit. Low and high warning and
failure  limits  are  programmable.  For
conventional  low  impedance  silver-sil-
ver  chloride  reference  electrodes  only
the  high  limits  are  useful.  For  high
impedance  reference  electrodes,  both
low and high limits are used.
Figure  8-2  shows  suggested  limits  for
low impedance reference electrodes.
Figure  8-3  shows  suggested  limits  for
high  impedance  glass  reference  elec-
trodes.
FIGURE 8-2. Suggested warning and Failure Limits for Low
Impedance Reference Electrodes
The  impedance  of  a  typical  silver-silver  chloride  reference  elec-
trode  is  less  than  40  kilohms.  If  the  impedance  is  greater  than
about  140  kilohms  the  reference  electrode  has  failed.  Failure  is
usually  caused  by  a  plugged  or  coated  reference  junction  or  a
depleted  electrolyte  fill  solution  (gel).  The  reference  impedance
will also be high if the sensor is out of the process liquid.
FIGURE 8-3. Suggested warning and Failure Limits for High
Impedance Glass Reference Electrodes.
The limits for a high impedance glass reference electrode are the
same as the limits for a high impedance glass measuring electrode.