Wiring the 9739 MVD transmitter to the sensor
Table 4-3
Sensor and transmitter terminal designations
continued
Wire color
Sensor terminal
Transmitter terminal
Function
Violet
7
7
Temperature +
Gray
8
8
Right pickoff −
White
9
9
Left pickoff −
b. Tighten the screws to hold the wires in place.
c. Ensure integrity of gaskets, grease all O-rings, then replace the junction-box and transmitter housing covers and
tighten all screws, as required.
Sensor and transmitter terminals
Figures 4-5 through 4-7 identify the terminals of the sensors that you can connect to the 9739 MVD transmitter, and
Figure 4-8 identifies the 9739 MVD transmitter terminals.
Figure 4-5
ELITE, H-Series, T-Series, and some F-Series sensor terminals
D
I
H
F
E
A B C
G
A
Violet
B
Yellow
C
Orange
D
Brown
E
White
F
Green
G
Red
H
Gray
I
Blue
18
Micro Motion 9739 MVD Transmitters