Справочник Пользователя для A&D MS-70

Скачать
Страница из 75
MS-70 / MX-50 / MF-50 / ML-50 
 
  The analyzer is equipped with a serial interface as standard. It can be connected to 
a printer or computer. 
 
 
  The analyzer can calibrate the weighing sensor (Use special mass.) and drying 
temperature (Use temperature calibrator for MS-70 and MX-50 except MF-50 and 
ML-50). The analyzer can output the data required at GLP, GMP and ISO at the end 
of the calibration. 
 
 
  The analyzer has a self check function that can detect function errors. 
 
 
  The analyzer displays the current change of moisture content per one minute [%/ min] 
in real-time. It can be used for the reference to find the analyzing mode. 
 
 
  The sample pans included in the standard accessory can be used repeatedly. 
 
 
  There is a test sample that is used to check the moisture accuracy. (The test sample 
is a part of standard accessory except ML-50) 
 
 
  The glass fiber sheets can be used to quick and precision measurement for liquid 
sample. (The glass fiber sheets are a part of standard accessory except ML-50) 
 
 
  A reference card is built into the bottom of the analyzer. 
 
 
Principle
Principle
Principle
Principle and Use
 and Use
 and Use
 and Use    
 
  The moisture analyzer, based on the principle of thermogravimetric analysis, dries a 
sample using a halogen lamp and obtains the moisture content in % and other 
results by the difference between the wet weight and dry weight.