Справочник Пользователя для Emerson 308014EN

Скачать
Страница из 166
2-9
Rosemount Tank Radar REX
Chapter 2 Safety
Installation Manual
308014EN, Edition 4
May 2013
2.3.7
Remote Display Unit 40 (RDU 40)
Figure 2-7. Approval label for the Remote Display Unit RDU40.
The following information is provided as part of the label of the Remote 
Display Unit 40 (RDU 40):
Name and address of the manufacturer 
(Rosemount Tank Radar AB).
CE Conformity Marking:
Year of construction
Marking for explosion protection:
Ex ib IIC T4 Gb (-40 °C 
Ta +70 °C)
Sira ATEX certificate number: Sira 00 ATEX 2062
The following instructions apply to equipment covered by certificate 
number Sira 00ATEX2062:
1
The equipment may be used with flammable gases and vapours 
with apparatus groups IIC, IIB and IIA and with temperature 
classes T1, T2, T3 and T4.
2
The equipment is only certified for use in ambient temperatures in 
the range -40 °C to +70 °C and should not be used outside this 
range.
3
Installation shall be carried out in accordance with the applicable 
code of practice.
4
Repair of this equipment shall be carried out in accordance with 
the applicable code of practice.
5
Certification marking as detailed in drawing number 9150 074-
980.