Section 5. Service and Troubleshooting
Incorrect configuration of 2160 Field 
Communication Unit (FCU)
• Check the Modbus communication address specified 
for the ATD device that represents the 2230 display in 
the 2160 FCU Slave Database. For the single tank 
version, the ATD address is equal to the Modbus 
address of the 2410 Tank Hub itself.
• Check configuration of communication parameters for 
the FCU Fieldbus ports 
• Check that the correct communication channel is 
selected
• See the Raptor System Configuration Manual 
(Document no. 300510EN) for more information on how 
to configure the 2160 FCU
Incorrect configuration of tank database in 
2410 Tank Hub
• Check the 2410 tank database; ensure that the 2230 
device is available and mapped to the right tank
• 2410 Tank Database; check that the ATD Modbus 
address is equal to the 2410 Temp Modbus address in 
the FCU Slave Database
• See the Raptor System Configuration Manual 
(Document no. 300510EN) for more information on how 
to configure the 2410 tank database
Connection to 2410 Tank Hub
• Check wiring to the 2410 Tank Hub
• Check the 2410 Tank Hub; check the Error LED or the 
integral display for information
Configuration of communication protocol
In TankMaster WinSetup:
• open the Protocols folder and check that the protocol 
channel is enabled
• check the protocol channel configuration (right-click 
the protocol channel MbMaster icon, choose the 
Properties option, and check port, parameters, 
modem)
The Status LED is blinking error 
codes
• Hardware errors
• Software errors
• See “Device Error Signals” on page 5-6
• Check Device Status information. See “Status 
Configuration can not be saved
Write protection switch is set to the ON 
position
Check write protection switch on the 2230.
Invalid measurement data (--.---)
Device failure
Check the devices connected to the Tankbus for possible 
hardware or software failure
Warning symbol appears in front 
of measurement value
Simulation mode active
Stop simulation mode in WinSetup (open WinSetup Set 
Simulation Mode
 window and click the Stop button)
Activity indicator shows a warning 
symbol
Communication failure
Check that the 2230 is configured in the 2410 tank 
database. See the Raptor System Configuration Manual 
(Document no. 300510EN) for more information on how to 
configure the 2410 tank database.
Nothing appears on the LCD 
display
• No power supply
• FISCO fuse broken
• Contrast settings
Check status LED (see “Device Error Signals” on 
page 5-6). 
If Status LED does not light:
• check power on Tankbus wiring
• check FISCO fuse
If Status LED lights:
• check contrast settings of the LCD display
Symptom
Possible cause
Action