10. Calibrate the Center Frequency DAC as follows:
a. At the
$
prompt, type:
calterm 114
and
press <ENTER>. (The
$
prompt will appear on
the screen when the calibration is complete.)
b. Record step completion on the Test Record.
11. Store the calibration data as follows:
a. At the
$
prompt, type:
calterm 787
and
press <ENTER>. (The
$
prompt will appear on
the screen when the calibration data has been
stored.)
b. Record step completion on the Test Record.
4-16
682XXB/683XXB MM
PRELIMINARY
CALIBRATION
CALIBRATION
CAUTION
When saving calibration data, turn-
ing off the instrument before the
$
prompt returns to the screen can
cause all stored data to be lost.