NI 5441 Specifications
38
ni.com
Calibration
Power
Specification
Value
Comments
Self-Calibration An onboard, 24-bit ADC and precision voltage reference are
used to calibrate the DC gain and offset. The self-calibration
is initiated by the user through the software and takes
approximately 75 seconds to complete.
is initiated by the user through the software and takes
approximately 75 seconds to complete.
—
External
Calibration
Calibration
The External Calibration calibrates the VCXO, voltage
reference, output impedance, DC gain, and offset.
Appropriate constants are stored in nonvolatile memory.
reference, output impedance, DC gain, and offset.
Appropriate constants are stored in nonvolatile memory.
—
Calibration
Interval
Interval
Specifications valid within 2 years of External Calibration.
—
Warm-up Time
15 minutes
—
Specification
Typical Operation
Overload Operation
Comments
+3.3 VDC
1.9 A
2.7 A
Typical.
Overload
operation occurs
when CH 0 is
shorted to
ground.
Overload
operation occurs
when CH 0 is
shorted to
ground.
+5 VDC
2.2 A
2.4 A
+12 VDC
0.46 A
0.5 A
–12 VDC
0.01 A
0.01 A
Total Power
22.9 W
27.0 W