Operating Basics
58
AFG3000 Series Quick Start User Manual
Generating a Differential Signal
A dual-channel instrument can be used to generate a differential signal by programming CH2 to output the complement of 
CH1. The CH2 settings can be configured conveniently by using the CH1 Complement function.
1. After setting the CH1 waveform param-
eters, push the front-panel Channel 
select 
 button to select CH2.
2. Push the front-panel Top Menu 
  
button, and then push the Output Menu 
bezel button.
3. Push the CH1 Complement bezel but-
ton.
4. The CH2 waveshape and timing param-
eters are copied from CH1, and the CH2 
amplitude settings are inverted from 
CH1.
NOTE. When CH1 is selected in the AFG310x or AFG325x series instrument, the instrument displays External Add Off/On 
in the Output menu.
CH2
CH1
1
3
2
4