Справочник Пользователя для Tektronix TDS460A
Command Descriptions
2–316
TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:SLEWRate:THReshold:LOW?
THReshold
TRIGger
:
:
PULse
MAIn
:
:
?
<Space>
<NR3>
LOW
SLEWRate
<NR3> is the threshold, in volts.
TRIGGER:MAIN:PULSE:SLEWRATE:THRESHOLD:LOW 50E-3
sets the lower limit of the pulse slew rate trigger to 50 mV.
sets the lower limit of the pulse slew rate trigger to 50 mV.
TRIGger:MAIn:PULse:SLEWRate:WHEn
TDS 500C, 600B, & 700C Only
Sets or queries whether to check for a slewing signal that is faster or slower than
the specified delta time. This is equivalent to selecting Trigger When from the
main Trigger’s Slew Rate menu and entering a value in the resulting side menu’s
Trig if Less Than or Trig if Greater Than items.
the specified delta time. This is equivalent to selecting Trigger When from the
main Trigger’s Slew Rate menu and entering a value in the resulting side menu’s
Trig if Less Than or Trig if Greater Than items.
Trigger
TRIGger:MAIn:PULse:SLEWRate:WHEn { FASTERthan | SLOWERthan }
TRIGger:MAIn:PULse:SLEWRate:WHEn?
FASTERthan
TRIGger
:
:
PULse
MAIn
:
:
WHEN
SLEWRate
SLOWERthan
<Space>
?
FASTERthan sets the trigger to occur when the slew is faster than the set
volts/second rate.
volts/second rate.
SLOWERthan sets the trigger to occur when the slew is slower than the set
volts/second rate.
volts/second rate.
Arguments
Examples
Group
Syntax
Arguments