Справочник Пользователя для Marathon Computer AACC 2000

Скачать
Страница из 103
Marathon Monitors Inc.
AACC 2000 Carbon                                                                       Nov.  1, 1997
84
Select ‘
YES’ with 
 or 
YES
Start calibration
Wait for calibration to
buSy
Busy calibrating
complete.
donE
PV input calibration completed
FAIL
Calibration failed
Note.   When a DC input module is installed for the first time, or there is a requirement to change one,
then the microprocessor in the controller needs to read the factory calibration data stored in the module.
Select ‘
FACt’ as the calibration value.   Step to ‘GO’ and start calibration.
DC Output Calibration
The following parameters apply to DC output modules ie for 
rcAL = 1A.Hi to 3A.Lo
cAL.H
Output Calibration High
0
0 = Factory set calibration.
Trim value until output = 9V, or
18mA
cAL.L
Output Calibration Low
0
0 = Factory set calibration.
Trim value until output = 1V, or
2mA
User calibration
UCAL
User calibration enable
Yes/no
pt1.L
Low calibration point for Input 1
The factory calibration point at which the
low point offset was performed.
pt1.H
High calibration point for Input 1
The factory calibration point at which the
high point offset was performed.
OF1.L
Offset Low for Input 1
Calculated offset, in display units.
OF1.H
Offset High for Input 1
Calculated offset, in display units.
pt2.L
Low calibration point for Input 2
The factory calibration point at which the
low point offset was performed.
pt2.H
High calibration point for Input 2
The factory calibration point at which the
high point offset was performed.
OF2.L
Offset Low for Input 2
Calculated offset, in display units.
OF2.H
Offset High for Input 2
Calculated offset, in display units.
Name
Description
Values
Meaning
PASS
PASS
Password configuration
ACC.P
FuLL or Edit level password
cnF.P
Configuration level password
Exit
Exit
Exit configuration
no/YES