USB-9211/9211A User Guide and Specifications
Figure 12.  Type K Errors
Figure 13.  Type R and S Errors
Cold-Junction Temperature Measurement Accuracy
Heat from other nearby heat sources can cause errors in thermocouple 
measurements by heating up the terminals so that they are at a different 
temperature than the cold-junction compensation sensor used to measure 
the cold junction. The thermal gradient generated across the terminals can 
cause the terminals of different channels to be at different temperatures, so 
the resulting measurement creates errors not only in absolute accuracy but 
also in the relative accuracy between channels.