Page 37                      SDS Data Logger Manual #4565 Rev D 
32. Survey Deviation Alarm Log 
 
The survey deviation alarm log will log all instances where the temperature of one of the survey channels 
goes above or below the deviation band. The alarm log is on a one-minute update, so any alarm would 
have to be present for more than a minute to be logged.  The start and stop time of the alarm is listed at 
the top of the screen on the line “AL 1 Start at”.  The screen will display the each channel and which 
channel was high or low.  This log will be cleared out when the next survey is run. 
34. Clear Logged Data 
 
This screen will allow the user to clear out the logged data.  This menu option is only accessible with the 
SSi Special pass code that can be obtained from Super Systems Inc support.  To begin clearing out the 
datalog data, enter a 1 and press Enter.  The middle-left of the screen will display “*** Busy ***” while 
the SDS instrument is clearing out the data.  When the instrument is finished clearing out the datalog 
data, the message will change to “Clearing complete.”  If there was an error, the SDS instrument will 
display “File error    -13”.  For SDS firmware versions 1.09 or lower (SDS menu 13 – 
Revision Display
), it 
is suggested that the datalog files be periodically deleted from the SDS data logger.  All survey data that 
is captured on the data logger is transferred to the PC using the SDS Reporter software and can be 
reviewed at any time from that PC.
  
Any data logger that retains datalog files for more than one hundred 
ten (110) days, or roughly three and a half (3 ½) months, will begin to experience issues with the data 
logger.  Datalog files can also be deleted by using the SDS 
→ Download Surveys and Data menu option 
on the SDS Reporter software to bring up the survey download screen.  From the Download screen, the