Super Systems Inc. 
Page 6                      SDS Data Logger Manual #4565 Rev D 
 
  
Degrees F. 
A/D 
Accuracy 
Accuracy  
Accuracy 
TC Type 
Min 
Max 
Full scale 
Full Scale 
At 1600 Degrees F. 
At 1600 Degrees F with 
thermister error 
B 32 
3308 
0.0124% 
0.0179% 
0.29 
0.65 
32 4208 0.0062% 
0.0067% 
0.11 
0.47 
-328 1832  0.0031% 
0.0032% 
0.05 
0.41 
-346 2192  0.0031% 
0.0036% 
0.06 
0.42 
-328 2502  0.0031% 
0.0045% 
0.07 
0.43 
-328 2372  0.0031% 
0.0052% 
0.08 
0.44 
NNM 
0 1409 0.0031%  0.0033% 
0.05 
0.41 
-58 3214  0.0062% 
0.0118% 
0.19 
0.66 
-58 3214  0.0124% 
0.0133% 
0.21 
0.57 
-328 752  0.0062% 
0.0119% 
0.19 
0.55 
*** Note: Accuracy at 1600 Degrees F with thermister error requires SDS firmware level 1.10 or later *** 
Software Installation 
You will find the SDS Recorder software CD that ships with the unit includes all the reporting and 
administrative utilities.  Insert the CD into the CD ROM device on your computer and click the setup.exe 
file. You will be given a step-by-step instruction for completing the installation process.  See 
Installing the 
SDS Software
 section for more details. 
Calibration 
The SDS data logger is calibrated prior to shipment.  The 
calibration process is available through menu option 22 on the 
data logger.  Instructions for calibration are included in this 
manual.  Super Systems calibrates the SDS data logger using 
NIST traceable instruments that are listed on the calibration 
certificate included with the device.  The default calibration is 
performed using a zero and span routine for the millivolt range of 
0 – 80mV.  Each range that is being used should be calibrated.  
For a list of suggested mV ranges by thermocouple types, please 
refer to the section on Input Calibration in the manual.  Other 
calibration ranges are available upon request. 
An input device will be used to input a specific type K 
thermocouple value and the data logger value will be recorded.  
This will be performed at different temperatures and printed with the certificate.