NI PXI/PCI-5114 Specifications
12
ni.com
TClk Specifications
National Instruments TClk synchronization method and the NI-TClk driver 
are used to align the sample clocks on any number of SMC-based modules 
in a chassis. For more information about TClk synchronization, refer to the 
NI-TClk Synchronization Help, which is located within the NI High-Speed 
Digitizers Help
.
Specifications are valid for any number of modules installed in one 
NI PXI-1042 chassis.
All parameters set to identical values for each SMC-based module.
Sample Clock set to
 
250 MS/s and all filters are disabled.
For other configurations, including multichassis systems, contact 
NI Technical Support at 
ni.com/support
.
Note
Although you can use NI-TClk to synchronize nonidentical modules, these 
specifications apply only to synchronizing identical modules.
Specification
Value
Comments
Intermodule SMC Synchronization Using NI-TClk for Identical Modules (Typical)
Skew
500 ps
Caused by clock and analog 
path delay differences
No manual adjustment 
performed
Skew After Manual 
Adjustment
<20 ps
For information about 
manual adjustment, refer 
to the Synchronization 
Repeatability Optimization
 
topic in the NI-TClk 
Synchronization Help

For additional help with the 
adjustment process, contact 
NI Technical Support at 
ni.com/support
.
Sample Clock 
Adjustment 
Resolution
<20 ps