NI 4070/4072 Calibration Procedure
21. Refer to Table 4 for the appropriate calibrator outputs and parameter 
values as you complete the following steps:
a.
On the calibrator, output the value listed in the Calibrator Output 
column in Table 4 for the current iteration.
b.
Call 
niDMM_ConfigureMeasurement
 with Mode set to 
NIDMM_VAL_AC_VOLTS
 and the remaining parameters as shown 
in Table 4 for the current iteration. 
c.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 16.
d.
Call 
niDMM_ConfigureMeasurement
 again, changing Mode to 
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
e.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 16.
22. Output 50 V at 30 Hz on the calibrator.
Table 4.  niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output
niDMM_ConfigureMeasurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
5 V
50 Hz
NIDMM_VAL_AC_VOLTS
5
5e–6
5 V
50 Hz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
5e–6
2
5 V
1 kHz
NIDMM_VAL_AC_VOLTS
5
5e–6
5 V
1 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
5e–6
3
5 V
1 kHz
NIDMM_VAL_AC_VOLTS
50
50e–6
5 V
1 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
50e–6
4
5 V 
1 kHz
NIDMM_VAL_AC_VOLTS
300
300e–6
5 V
1 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
300e–6
5
5 V
20 kHz
NIDMM_VAL_AC_VOLTS
5
5e–6
5 V
20 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
5e–6
6
5 V 
50 kHz
NIDMM_VAL_AC_VOLTS
5
5e–6
5 V 
50 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
5e–6
7
5 V
100 kHz
NIDMM_VAL_AC_VOLTS
5
5e–6
5 V
100 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
5e–6
8
5 V
300 kHz
NIDMM_VAL_AC_VOLTS
5
5e–6
5 V
300 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
5e–6