Справочник Пользователя для National Instruments SCC-AI07

Скачать
Страница из 18
SCC-AI Series Isolated Analog Input Modules
10
ni.com
Table 1-2 shows suggested values for some of the parameters. Use default 
values for parameters not shown.
9.
Click Test to test the channel. You should be able to view your 
measurement in the window that appears.
You have finished configuring the NI-DAQmx task or global channel.
Creating a Virtual Channel in Traditional NI-DAQ
To create a Traditional NI-DAQ virtual channel, complete the following 
steps.
1.
Open Measurement & Automation Explorer.
2.
Right-click Data Neighborhood and select Create New.
3.
Select Traditional NI-DAQ Virtual Channel and click Finish.
4.
Click Next to select Analog Input as the type of channel to configure.
5.
Select Voltage from the drop-down list and click Next.
6.
Enter the rest of the values as they apply to your setup and click Next.
7.
On the next screen, specify information about the E Series DAQ device 
being used and click Finish.
8.
To test the channel, click Test.
You have finished configuring the Traditional NI-DAQ virtual channel.
Table 1-2.  SCC-AIXX Suggested Values
Module
Input Range
SCC-AI01
±42 V
SCC-AI02
±20 V
SCC-AI03
±10 V
SCC-AI04
±5 V
SCC-AI05
±1 V
SCC-AI06
±100 mV
SCC-AI07
±50 mV
SCC-AI13
±10 V
SCC-AI14
±5 V