This category refers to local-level electrical distribution, such as 
that provided by a standard wall outlet, for example, 115 V for U.S. 
or 230 V for Europe.
Caution
Do not connect the NI 9472/9474 with screw 
terminal to signals or use for measurements within 
Measurement Categories III or IV.
NI 9472 with DSUB Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-COM .............................. 30 VDC max
Isolation
Channel-to-channel .................... None
Channel-to-earth ground
Continuous ........................... 60 VDC, 
Measurement Category I
Withstand ............................. 1,000 V
rms
, verified by a 5 s 
dielectric withstand test 
Measurement Category I is for measurements performed on 
circuits not directly connected to the electrical distribution system 
referred to as MAINS voltage. MAINS is a hazardous live electrical