NI 9235/9236 
4
ni.com
Caution
Substitution of components may impair 
suitability for Class I, Division 2.
Caution
For Zone 2 applications, install the system 
in an enclosure rated to at least IP 54 as defined by 
IEC 60529 and EN 60529.
Caution
For Zone 2 applications, connected signals must 
be within the following limit:
Capacitance .......................... 0.2 
μF max
Special Conditions for Hazardous Locations Use in Europe
This equipment has been evaluated as Ex nA IIC T4 equipment 
under DEMKO Certificate No. 07 ATEX 0626664X. Each module 
is marked 
 II 3G and is suitable for use in Zone 2 hazardous 
locations. If you are using the NI 9235/9236 in Gas Group IIC 
hazardous locations or in ambient temperatures of 
–40 °C
≤ Ta ≤ 70 °C, you must use the device in an NI chassis 
that has been evaluated as EEx nC IIC T4, Ex nA IIC T4, or 
Ex nL IIC T4 equipment.