NI 9235/9236 
You can connect a quarter-bridge sensor to each channel. Each 
channel has an EXC terminal that provides the excitation voltage 
stimulus, an AI terminal that measures the bridge voltage, and an 
RC terminal that provides the quarter-bridge completion. Refer to 
Figure 2 for an illustration of how to connect quarter-bridge 
sensors to the NI 9235/9236.
Figure 2.  Quarter-Bridge Connections for the NI 9235/9236
*
For best system accuracy, set up the connections to EXC and RC with 
equal lengths of an identical wire type and gauge.
**
 For best system accuracy, set up the connection to AI directly at the 
sensor instead of shorting AI to RC directly at the terminals.
EXC*
AI**
RC*
NI 9235/9236