SCC-CO20 Isolated Current Output Module User Guide
8
ni.com
Figure 5.  Connecting the SCC-CO20 to a Grounded External Loop Supply 
and a Grounded Load
For information about how to configure the SCC-CO20 module using 
NI-DAQmx, refer to the SCC Quick Start Guide.
Calibrating Gain and Offset Errors
The SCC-CO20 is calibrated at the factory before shipment. If you want 
to adjust the gain and offset of the SCC-CO20 in your system using an 
E/M Series DAQ device, you need a current meter (ammeter) with an 
accuracy of at least ±2
μA.
To adjust the gain and offset of the SCC-CO20, complete the following 
steps:
1.
Disconnect the load from the SCC-CO20 you want to calibrate.
2.
Set an output range of 0 to 10 V for AO (X), the E/M Series DAQ 
device channel connected to the SCC-CO20.
3.
Connect the negative lead of the ammeter to the current sink (pin 2) 
and the positive lead to the current supply (pin 3) of the SCC-CO20.
4.
Using your software, have AO (X) output 10 mVDC.
5.
Adjust the offset potentiometer on the top of the SCC-CO20 until the 
ammeter reads 20 
μA. The offset is now calibrated.
6.
Using your software, have AO (X) output 9 VDC.
7.
Adjust the gain potentiometer on the top of the SCC-CO20 until the 
ammeter reads 18 mA. The gain is now calibrated.
Load 
 
0–20 mA 
SINK 
SCC-CO20 
ISO GND 
–