NI 9144 User Guide and Specifications
NI 9233
As a DSA module, the NI 9233 does not synchronize to other modules and
free-runs at its own fixed rate.
free-runs at its own fixed rate.
NI 9233 Configure ADC
The NI 9233 (and NI 9229/9239) converts at various rates, controlled by
the field in the ADC conversion command.
the field in the ADC conversion command.
0x65
Data Byte
0xA0
CRC value
Table 42. NI 9233 Vendor Configuration Extensions
Index
Sub
Type
R/W
Description
0x2002
0
U32
R/W
Configure ADC,
default = 0x0A
default = 0x0A
0x2100
0
ARR:U32
—
Calibration = 16
1
R
Ch0 Offset
2
R
Ch0 Gain
3
R
Ch1 Offset
...
...
...
8
R
Ch3 Gain
9
R
External Ch0
Offset
Offset
...
...
...
Table 43. NI 9233 Scan List Format
Bits
Field
7
Turbo Disable (NI 9233 only)
6:2
Clock Divisor
1:0
Clock Source = 2
Table 41. NI 9219 Configuration 1: Command Bytes (Continued)
Command Byte Value