Chapter 3
Hardware Overview
© National Instruments Corporation
3-7
 
Figure 3-7.  Inside-Region Analog Triggering Mode
In high-hysteresis analog triggering mode, the trigger is generated when the 
signal value is greater than highValue, with the hysteresis specified by 
lowValue, as shown in Figure 3-8.
Figure 3-8.  High-Hysteresis Analog Triggering Mode
In low-hysteresis analog triggering mode, the trigger is generated when the 
signal value is less than lowValue, with the hysteresis specified by 
highValue, as shown in Figure 3-9.
Figure 3-9.  Low-Hysteresis Analog Triggering Mode
highValue
Trigger
lowValue
highValue
Trigger
lowValue
highValue
Trigger
lowValue