USB-9201/9221 User Guide and Specifications
Safety Guidelines
Operate the USB-9201/9221 only as described in these operating 
instructions.
Caution
Although the NI 9201/9221 might have more stringent certification standards 
than the USB-9201/9221, when used with the USB-9162 carrier, the combined system 
might be limited. Refer to the 
 section for more information.
Hot Surface
This icon denotes that the component may be hot. Touching this component 
may result in bodily injury.
Caution
Do not disconnect I/O-side wires or connectors unless power has been switched 
off or the area is known to be nonhazardous.
Caution
Do not remove modules unless power has been switched off and the area is known 
to be nonhazardous.
Caution
The USB-9201/9221 is not certified for use in hazardous locations.
Safety Guidelines for Hazardous Voltages
You can connect hazardous voltages to the USB-9201/9221 with the screw 
terminal only. Do not connect hazardous voltages to the USB-9201/9221 
with DSUB.
If hazardous voltages are connected to the module, take the following 
precautions. A hazardous voltage is a voltage >42.4 V
pk
 or 60 VDC to earth 
ground.
Caution
Ensure that hazardous voltage wiring is performed only by qualified personnel 
adhering to local electrical standards.
Caution
Do not mix hazardous voltage circuits and human-accessible circuits on the same 
module.
Caution
Make sure that devices and circuits connected to the module are properly 
insulated from human contact
Caution
When module terminals are live with hazardous voltages, make sure that the 
terminals are not accessible by using the high voltage screw terminal backshell. Refer to 
the 