Physical Characteristics
If you need to clean the module, wipe it with a dry towel.
Dimensions
With combicon................................ 14.0 cm 
× 8.8 cm × 2.5 cm
........................................................ (5.52 in. 
× 3.47 in. × 1.00 in.)
25-pin DSUB .................................. 12.6 cm 
× 8.8 cm × 2.5 cm
........................................................ (4.95 in. 
× 3.47 in. × 1.00 in.)
Screw-terminal wiring............................ 12 to 24 AWG copper conductor 
wire with 10 mm (0.39 in.) of 
insulation stripped from the end
Torque for screw terminals .................... 0.5 – 0.6 N · m (4.4 – 5.3 lb · in.)
Weight
USB-9201/9221 with combicon ..... Approx. 250 g (8.8 oz)
USB-9201/9221 with DSUB .......... Approx. 245 g (8.6 oz)
Safety
Maximum Voltage
Connect only voltages that are within these limits.
AI to COM
............................................. ±60 VDC
Isolation Voltages for USB-9201/9221 with Screw 
Terminals
Channel-to-channel isolation ................. None
Channel-to-earth ground isolation
Continuous ...................................... 250 V
rms
, Measurement 
Category II
Withstand ........................................ 2,300 V
rms
, verified by a 
dielectric withstand test
Measurement Category II is for measurements performed on circuits 
directly connected to the electrical distribution system. This category refers 
to local-level electrical distribution, such as that provided by a standard 
wall outlet (for example, 115 V for U.S. or 230 V for Europe). Do not use 
this module with Measurement Category III or IV voltages.