Chapter 6
Calibrating Images
6-2
ni.com
Refer to Chapter 5, 
, for more 
information about applying calibration information before making 
measurements.
Defining a Calibration Template
You can define a calibration template by supplying an image of a grid or 
providing a list of pixel coordinates and their corresponding real-world 
coordinates. This section discusses the grid method in detail.
A calibration template is a user-defined grid of circular dots. As shown in 
Figure 6-1, the grid has constant spacings in the x and y directions. You can 
use any grid, but follow these guidelines for best results: 
The displacement in the x and y directions should be equal (dx = dy).
The dots should cover the entire desired working area.
The radius of the dots in the acquired image should be 6–10 pixels.
The center-to-center distance between dots in the acquired image 
should range from 18 to 32 pixels, as shown in Figure 6-1.
The minimum distance between the edges of the dots in the acquired 
image should be 6 pixels, as shown in Figure 6-1.
Figure 6-1.  Defining a Calibration Grid
Note
Click Start»Programs»National Instruments»Vision»Documentation» 
Calibration Grid to use the calibration grid installed with IMAQ Vision. The dots have 
radii of 2 mm and center-to-center distances of 1 cm. Depending on your printer, these 
measurements may change by a fraction of a millimeter. You can purchase highly accurate 
calibration grids from optics suppliers, such as Edmund Industrial Optics.