Справочник Пользователя для A&D AD-4212B-101/102/201/301
47
11-2 GLP Report
Set the following parameters to output the report.
To print the report, set the “GLP output ( info )” parameter to “1” and use MODE 3 of the
AD-8121B. For details on using the printer, refer to “17-1 Connection to the AD-8121B Printer”.
To output the report to a personal computer using the RS-232C interface, set the “GLP output
( info )” parameter to “2”.
If the time and date are not correct, set the correct time and date in “Clock ( Cl adj )” of the
function table.
Note
For operational details about calibration and calibration test, refer to “8. CALIBRATION”.
For operational details about calibration and calibration test, refer to “8. CALIBRATION”.
Calibration report
When the setting is “info 1”:
When the setting is “info 2”:
AD-8121 format
General data format
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration
weight
Signature
CALIB RATED(EX T.)
CAL.W EIGHT
CAL.W EIGHT
1234
+ 50.00000 g
SIGNA TURE
- - - - - - - -
A & D
S/N
ID
DATE
TIME
ID
DATE
TIME
AD4212B-102
01234567
ABCDEFG
2000/12/31
12:34:56
A & D<TERM>
AD 42 12 B-1 02 <TERM>
S/N 0 12 345 67 <TERM>
ID AB CDE FG <TERM>
DAT E<TERM>
200 0/ 12 /3 1<TERM>
TIM E<TERM>
1 2: 34: 56 <TERM>
CAL IB RA TE D( EXT .)<TERM>
CAL.WEIGHT
+ 50 .0 00 00 g <TERM>
SIG NA TU RE <TERM>
<TERM>
<TERM>
-- - - -- - - -- - - - -- - <TERM>
<TERM>
<TERM>
AD 42 12 B-1 02 <TERM>
S/N 0 12 345 67 <TERM>
ID AB CDE FG <TERM>
DAT E<TERM>
200 0/ 12 /3 1<TERM>
TIM E<TERM>
1 2: 34: 56 <TERM>
CAL IB RA TE D( EXT .)<TERM>
CAL.WEIGHT
+ 50 .0 00 00 g <TERM>
SIG NA TU RE <TERM>
<TERM>
<TERM>
-- - - -- - - -- - - - -- - <TERM>
<TERM>
<TERM>
<TERM>
CR LF
CR
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
CR
LF