14. DATA MEMORY 
Data memory is a function to store weighing data or calibration data to display or output later. The 
data memory function can also store multiple upper/lower limit values or unit mass values, for later 
selection as necessary.   
One of the following data sets can be stored:   
 
 Weighing data (Up to 200 sets. 100 sets when the time and date are added.)  
 
 Calibration and calibration test data (latest 50 sets) 
 
 Unit mass in the counting mode (Up to 20 sets) 
 
 Upper/lower limit values (Up to 20 sets) 
Note 
Data memory function is not available when the BCD output (OP-01) is installed. 
 
14-1  Notes on Using Data Memory 
To use the memory function, set the “Data memory ( data )” parameter of the function table. In 
addition, for weighing data, set the “Time/Date output (5-td)” parameter. For details on setting the 
data memory, refer to “10. FUNCTION TABLE”. 
For weighing data, the data contents to be stored and the storage capacity depend on the “Time/Date 
output (5-td)” parameter setting. 
 
Releasing “Clr” 
If a different type of data exists in memory when the data is stored, “Clr” appears blinking in the 
upper left of the display. For example, you want to store weighing data but calibration data or unit 
mass data remains in memory.   
 
Under such a condition, before storing data, delete the data in memory as follows: 
1  Press and hold the PRINT key until SClr no  with “no” blinking is displayed, then release the key. 
2  Press the RE-ZERO key to display SClr go  with “go” blinking.   
  The type of data stored in memory appears on the upper left of the display as shown below: 
Unit mass in the counting mode 
pC 
Weighing data without time and date 
-d- 
Weighing data with time and date 
d-t 
Calibration or calibration test data 
Hi5 
Upper/lower limit values 
Cp 
3  Press the PRINT key to delete all the data in memory. 
4 The balance displays  end    and returns to the weighing mode.