11
3-2  When Building into a System 
The AD-4212A/B is a precision balance. When it is built into a system and used, errors such as 
unstable weight values may occur due to static electricity, vibration and materials used for the 
devices near the balance. 
When using the balance that is built into a system, take the following precautions. 
 
  Errors due to a static charge 
   When the ambient humidity is less than 45% RH, insulators such as plastic or glass are prone to 
static electricity. When charged material comes close to the balance, a pulling force is generated 
between the charged material and the weighing pan. This causes an unstable weight value. 
   To protect the balance against a discharge generated by charged material when it comes 
close to the balance, make sure to earth ground the weighing unit and the display unit. 
   (Static electricity generated by static induction will not be canceled by earth-grounding.) 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
      
Measures to take (Plastic is used in the measures below. They can be applied to glass, too.) 
  When the sample or devices are plastic 
   Use a static eliminator that generates no air blow such as the 
AD-1683, DC static eliminator, to remove static electricity. 
  Place the sample in a container that is made of a conductive 
material such as metal and that can be sealed and weigh it. 
 
  When the sample is powdery 
  When the balance is used in combination with a feeder for 
batch weighing of powdery samples, samples may be 
charged by rubbing sample particles against each other. 
  Use a static eliminator and perform weighing while removing 
static electricity. 
 
  When the sample container is made of material that is prone to static electricity such as plastic 
  Cover the outside of the container with a metal such as aluminum foil. 
  Apply an anti-static agent onto the container. 
 
  When making a breeze break using plastic 
  Apply an anti-static agent onto the breeze break. 
  Use a conductive acrylic fiber. 
 
  When plastic exists in the balance installation site 
  Cover the plastic with a grounded metal. 
  Apply an anti-static agent onto plastic. 
Metal case
Charged
material
Grounding
A pulling force generated between
the charged sample and the balance
causes an unstable weight value.
Charged plastic breeze break
The pulling force
makes the weight
value unstable