Справочник Пользователя для Zebra Technologies 105SL
38
105SL User Guide
6/4/2004 11342L Rev. 8
Printer Operation
Adjust and Calibrate Sensors
Positioning the Media (Transmissive) Sensors
The correct positioning of the media sensors is important—it can make the difference between
a perfect label and a call to Technical Support!
a perfect label and a call to Technical Support!
The web or gap sensor, also known as the “transmissive sensor,” detects the gap between
labels. The transmissive sensor actually consists of two sections: a light source (the lower
media sensor) and a light sensor (the upper media sensor). The media passes between the two.
labels. The transmissive sensor actually consists of two sections: a light source (the lower
media sensor) and a light sensor (the upper media sensor). The media passes between the two.
The upper media sensor must be positioned:
• Directly over the hole or notch, or
• Anywhere along the width of the media if there is a gap between labels.
• Anywhere along the width of the media if there is a gap between labels.
To adjust the upper and lower media sensors, complete these steps:
1.
Remove the ribbon (if it is installed). Refer to
2.
Locate the upper media sensor (
). The upper media sensor “eye” is directly
below the adjustment screw head.
3.
Slightly loosen the upper media sensor adjustment screw (use a Phillips-head
screwdriver).
screwdriver).
4.
Using the tip of the screwdriver, slide the upper sensor along the slot to the desired
position.
position.
5.
Tighten the upper media sensor adjustment screw.
Figure 30 • Upper Media Sensor Adjustment
Note •
If you are using continuous media, position the upper media sensor over the media
with the lower media sensor directly below it so that the printer can detect an out-of-paper
condition.
condition.