Ultrasound Output Connector
The ultrasound output connector used with all standard 
generators is a high voltage (5000V) coaxial style SHV-
BNC connector. This connector provides superior shielding 
of electrical noise, compared to other types of connectors. 
The ultrasound output connector mates with fully shielded 
coaxial ultrasound cables that are secured with a simple and 
reliable quarter-turn bayonet style attachment mechanism. 
Figure 3-13 
Ultrasound Output Connector
CAUTION
The ultrasonic output from 
this connector (that drives 
the attached ultrasonic load) 
is a very high AC voltage 
(1200VAC).  At high power lev-
els this can exceed 2 amperes 
of current and must be securely terminated 
via the ultrasound cable for safe operation. 
Use original equipment ultrasound cables 
for safe and reliable system operation. Im-
properly assembled ultrasound cables can 
result in high voltage arcing and will destroy 
the ultrasound connectors.
Do not use your generator if there is any 
evidence of arcing (black carbon deposits) 
on either the ultrasound output connector 
or the ultrasound cable connectors.
Configuration Port Connector
The configuration port connector is a DB-9M (standard 
D-subminiature 9 circuit male) typically used for RS-232 
serial communications. This serial port (DTE) connects to a 
serial port (DCE) on a computer via a standard 9 pin serial 
cable. If the computer does not have a serial port, you may 
use a USB-to-serial conversion cable.
This port is used for field updates to the generator firmware, 
without removing the enclosure cover. This port can also be 
used with a software application running on a Windows PC 
to modify the factory default system settings and hardware 
configurations. Contact your local sales representative for 
software availability information and access to documenta-
tion that will allow you to make use of the configuration 
port features.