Density (Consistency) Meter     LQ500
  
Introduction 
The LQ500 density (consistency) meter uses 
microwave phase shift technology to determine 
concentrations of solids in the material to be measured 
flowing through pipes. It can perform a stable and 
realtime density (consistency) measurement because 
this technology is not affected by flow velocity along 
with fluid color, and also is not easily affected by 
contaminants and low process pressure rate. As the 
LQ500 has no moving parts, it is reliable and virtually 
maintenance free. 
Since the output of the LQ500 is theoretically linear, it 
can be applied to a wide range of density (consistency) 
measurement.
 
<Notice> 
The LQ500 requires a full pipe to measure the 
density (consistency). Contact Toshiba before 
installation in the following cases: 
<Possibility of unfilled condition> 
(a) When it is installed at the discharge of a pump.
(b) When installation is horizontal, and unfilled 
condition occurs inside the pipe. 
(c) A process where the pipe becomes unfilled 
when the operation is stopped. 
 
 Cable for power supply 
(Detector/converter) 
Ground resistance
100 ohm or less 
Density (consistency)
measurement output 4-20mA
Cable for communication
(Detector/converter)
AC power supply
DI / DO 
External Sync control
contact input 
 
Figure
 
1. LQ500
 
Configuration
 
Diagram 
■ 
Standard Configuration 
• Density (Consistency) Meter: 
1 set 
(Detector and converter separate mounted) 
• Accessories: 
1 set (see Table 1 below)
 
Table 1. Standard Accessories 
Items Specifications 
Quantity
Power supply 
cable 
Between detector and 
converter (*1) 
10 m 
(32.8 ft) 
Communication 
cable 
Between detector and 
converter (*1) 
10 m 
(32.8 ft) 
Fuse 
2A(T), 250 V 
(glass tube, 5.2 dia. x 20 mm)
Document Instruction 
manual  1 
Note 1: Need to prepare a power supply cable for the 
LQ500. 
Refer to the section of cable specifications at the 
overall specifications in detail. 
 
 
     
 
 
Figure 2. LQ500 Density (consistency) Meter 
   
 
Specifications 
■ 
Overall Specifications 
Measurement method: 
Microwave
 
phase
 
difference method 
Measurement
 
range: 
Meter size 
50 mm (2”) 
80 to 300 mm
(3” to 12”) 
Span (*2) 
2 to 50 %TS (*1) 1 to 50%TS (*1)
Lower limit setting 
range (4 mA) 
0 to 48%TS 
0 to 49%TS 
Upper limit setting 
range (20 mA) 
2 to 50%TS 
1 to 50%TS 
Setting increments 
0.1%TS 
*1   TS: Total Solids 
*2   Span = Upper range – Lower range 
*3   The material to be measured must be fluid 
and be filled evenly with no voids. 
Repeatability:
 
Meter size 
50 mm (2”) 
80 to 300 mm 
(3” to 12”) 
Repeatability ±0.02%TS ±0.01%TS 
Note 1 : Above values are the results of commuting  ability 
in the phase measurements of the converter. 
Note 2 : Density (consistency) determination 
repeatability for sample reagent; 
Meter size 
50 mm
(2”) 
80 to 300 mm
(3” to 12”) 
For the full 
scale value of 
2%TS or 
greater 
±2%FS ±2%FS 
Density 
(consistency)
determination 
repeatability 
For the full 
scale value of 
less than 2%TS 
 ±4%FS 
*The characteristics of sample reagent has errors due 
to sample tests such as uneven density (consistency) 
distribution. 
*Full scale is the maximum value in the measurement