i600 Series Scanners 
Message listing
Following is an alphabetic list of messages and corrective actions you 
can take if one of the following messages is encountered.
Messages
ID#
Operator Message
Background accessory changed 
while powered up
381
The Background Accessory was changed while the scanner was 
powered on. 
• Turn the power off to the scanner, wait a few seconds and 
power the scanner back on.
• If you get this message and you did not change the 
Background Accessory, clean the scanner.
Cable unplugged
319
Some device on the FireWire bus was unplugged/plugged in. 
• Check that all FireWire connectors are securely fastened.
Calibration failed
106, 331  • Clean the imaging guides. See Chapter 5, Maintenance for 
procedures. 
• Try to calibrate the scanner again. Be sure you are using the 
correct calibration target. See Chapter 3, “Calibrating the 
scanner”. 
• The calibration target is too small. A 12- x 12-inch calibration 
target is required for image calibration. 
• Calibration target is too narrow.
• If the problem persists, call Service.
Calibration succeeded 
105, 342
Informational message. The scanner has been successfully 
calibrated.
Call Service
Many ID 
numbers
A condition was detected that may require repair.
• Turn the power off to the scanner, wait a few seconds and 
power the scanner back on.
• If the problem persists, call Service and provide the ID number 
of the error.
Cannot calibrate with white 
background
184
Both a front and rear print head is installed in the scanner. Only 
one print head can be installed at one time. 
• Remove one of the print heads.
Cannot run with both printers 
installed
90
The White Background Accessory is installed in the scanner 
while the scanner is trying the calibrate.
• Remove the White Background Accessory before calibrating 
the scanner. 
Check document preparation
123, 135, 
81
• The gap between your documents is too small.
• Make sure the leading edges of the documents are aligned. 
See Chapter 3, “Document preparation” for more information.
• Check the condition of your feed module and separation roller 
tires and your pre-separation pad. Clean and replace if 
necessary.
Check imaging guides and lamps
149, 178, 
139
• Verify that the imaging guides are installed correctly. See 
Chapter 5, Maintenance.
• Clean both sides of the imaging guides and lower glass plate 
between the lamps.
• Check that the white patch on the imaging guides is clean and 
in good condition.
• If the problem persists, call Service.
Clean lower imaging guide
294
The lower imaging guide is dirty or there may be a small piece of 
paper in the imaging guide area. 
• Clean the imaging guides. See Chapter 5, Maintenance for 
procedures.