Справочник Пользователя для 3M #3968

Скачать
Страница из 58
3M™ Rapid Detection Reader 
3M Rapid Detection Reader (MODULES) 
3M Rapid Detection Reader (MODULES) 
Figure 2.  3M Rapid Detection Reader Features - Front View  
 
Turning on the 3M Rapid Detection Reader  
The power On/Off switch is located on the rear panel of the CM 
(Figure 1).  When the power is switched on, the 3M Rapid Detection 
Reader performs a number of internal self-tests, including IQC, to 
ensure the system is operating within specifications.  For additional 
information, refer to Running Quality Control section. 
When switched ON, the LEDs on the front of the TM illuminate blue.  
When the power-on sequence is complete, the Home screen is 
displayed (Figure 3).  The power-on sequence takes approximately  
1 minute.   
Once the Home screen is displayed, selections can be made by 
touching the appropriate prompt on the display (i.e., Run Assay to 
initiate an assay, or Menu to go to the Main Menu screen (Figure 4)).  
The display is touch-sensitive and can accept input from a bare/gloved 
finger or a blunt stylus.  
The 3M Rapid Detection Reader can be operated immediately after 
set-up using the pre-programmed default settings.  The date, local 
time and some settings can be edited manually through the Settings 
menu on the CM touch screen.  Most settings can be edited using the 
Reader Configuration Utility (RCU) supplied with the CM.  For 
additional information on editing user settings, refer to Managing 
Settings and Editing Settings Using the Reader Configuration Utility 
(RCU). 
Figure 3.  Home 
 
 
 
 
CAUTION.   
Pens or other sharp instruments will damage the touch screen. 
Touch 
Screen 
Display 
Lot Card 
Insertion Slot 
Port LEDs 
Test Cartridge 
Ports 
Control Module
Test Module (3 max)