3M™ Rapid Detection Reader 
Warnings and Precautions 
Summary of Warnings and Precautions 
The 3M Rapid Detection Reader and its related devices and accessories are designed to provide 
safe and reliable service when used according to the instructions provided in this Operator 
Manual. 
The following warnings and precautions should be followed in order to avoid unsafe actions with 
the 3M Rapid Detection Reader that could potentially result in personal injury or device damage. 
 
 
 
WARNING! 
To reduce the risk associated with hazardous voltage: 
• 
Unplug the 3M Rapid Detection Reader before cleaning. 
• 
Do not disassemble the 3M Rapid Detection Reader or any of its related 
components and accessories.  The instrument contains no operator-serviceable 
components. 
• 
Do not immerse any of the 3M Rapid Detection Reader components in any liquid. 
• 
Do not use the 3M Rapid Detection Reader if there is apparent damage to a power 
cord or power supply. 
To reduce the risk associated with potentially infectious patient samples: 
• 
Do not spill specimen or sample fluids on any of the 3M Rapid Detection Reader 
components or on the outside of the Test Cartridge. 
• 
If a spill occurs, disinfect external surfaces only, using a soft cloth containing a 
solution of 0.5% bleach, 70% isopropyl alcohol or 70% ethanol.  
 
 
CAUTION. 
To reduce the risk associated with incorrect results: 
• 
3M Rapid Detection Reader should only be used by trained professionals and 
operated in accordance with facility policies and procedures.  
• 
3M Rapid Detection Reader should only be operated in ambient temperature 
conditions between 15-30°C (59-86°F) and out of direct sunlight. 
To reduce the risk of instrument or accessory damage: 
• 
3M Rapid Detection Reader is not designed to withstand moisture, temperature 
extremes, severe shock or vibration. 
• 
3M Rapid Detection Reader is intended for use indoors, located on a stable, 
stationary surface such as a counter top. 
• 
Do not open any 3M Rapid Detection Reader device enclosures, this will void the 
warranty. 
To reduce the risk associated with environmental contamination: 
• 
Dispose of instrument and accessories in accordance with federal, state, and local 
requirements.