Справочник Пользователя для Honeywell SC3004

Скачать
Страница из 174
 page 118      008-0608-00
If the 
CALIBRATION TYPE
 is 5-Point Known Load Calibration...
The display will read 
DOING 5POINT CAL
, and prompt you to 
APPLY 00000 UNIT
 
(where 00000 and UNIT are previously entered as
 KNOWN POINT 1/5
 and 
DISPLAY. 
UNITS
).  When you have applied this load to the transducer, press [ENTER]. 
The display will read 
WORKING
, then 
APPLY 00250 UNIT
 (where 00250 and UNIT are 
previously entered as
 KNOWN POINT 2/5
 and 
DISPLAY. UNITS
). When you have 
applied this load to the transducer, press [ENTER].
The display will read 
WORKING
, then 
APPLY 00500 UNIT
 (where 00500 and UNIT are 
previously entered as
 KNOWN POINT 3/5
 and 
DISPLAY. UNITS
). When you have 
applied this load to the transducer, press [ENTER].
The display will read 
WORKING
, then 
APPLY 00750 UNIT
 (where 00750 and UNIT are 
previously entered as
 KNOWN POINT 4/5
 and 
DISPLAY. UNITS
). When you have 
applied this load to the transducer, press [ENTER].
The display will read 
WORKING
, then 
APPLY 01000 UNIT
 (where 01000 and UNIT are 
previously entered as
 KNOWN POINT 5/5
 and 
DISPLAY. UNITS
). When you have 
applied this load to the transducer, press [ENTER].
The display will now read 
RETURN PNTS.=NO
. If you press [ENTER], the display 
will indicate 
DONE
 and the instrument will return to the RUN mode.  If you use 
the [UP] button and [ENTER] to select 
RETURN POINTS=YES
, you may now apply 
the same loads again but in reverse order.  This will allow the instrument to 
measure and partially eliminate any hysteresis that may be present in the 
transducer.  After the loads have been applied again, the display will indicate 
DONE
 and the instrument will return to the RUN mode.